Remarkably Well-Designed and Incredibly Convenient
Wisdom-P is designed to facilitate a wide range of procedures through two detectors for clinics or hospitals.
Its compact and clever design delivers high-performance as well as cost-effective
operation through ease-of-use which fully meet the demands on quality and versatility with affordable price.
The overhead tube crane is basically manual, but it includes motorized movement for the z-axis to be synchronized with other radiographic components.
Configuration
Overhead Tube Crane
Wall Bucky Stand
4-Way or 6-Way Patient Table
Ceiling-Mounted Tube Support | |
---|---|
Display with color LCD screen on the tube housing | |
Workflow is enhanced through direct manipulation of examination parameters listed below | |
- SID (Source-image distance) measurement | |
- Display angle of tube rotation | |
- Display of the workplace: Bucky wall stand or table | |
- Display auto or manual mode | |
Tube Travel Range |
- Vertical travel motorized in auto mode - Vertical travel range: 150cm(59 in.) - Rotation range around the vertical axis: -150° to +180° manual Detent at 0°; ±90° - Rotation range around the horizontal axis: ±150°manual Detent at 0°; ±90° |
Horizontal Travel Range | Longitudinal: 332 cm (130.7 in.) with 4 m trolley Transverse: 225 cm (88.6 in.) with 3 m trolley |
Power Requirement |
UL: 120 VAC, 50/60 Hz, 500 VA CE: 220 ~ 240 VAC, 50/60 Hz, 500 VA |
6-Way Patient Table (Manual) | |
---|---|
The table can be equipped with cassettes or a mobile flat detector for optimal coverage and high patient throughput | |
Tabletop Dimensions | 220 cm x 80 cm (86.6 in. x 31.7 in.) |
Table Height | 54.5 cm to 84.5 cm; total lift 30 cm (tabletop) |
Tabletop Travel Range | Longitudinal: ±43 cm (±16.9 in.) Transverse: ±12 cm (±4.7 in.) |
Detector Travel Range | 44 cm (17.3 in.) |
Max. Patient Weight | 250 kg (551 Ib) |
Kick Switch | For table height and float adjustments (Wireless Foot Switch) |
Auto-tracking during table height adjustment (Only with ceiling-mounted tube support) |
X-ray tube follows table height adjustment; SID (source-image distance) is maintained |
Power Requirement | 100 ~ 240 VAC, 50/60 Hz, 3A |
Wall Bucky Stand (Manual) | |
---|---|
The stand can be equipped with cassettes or a mobile flat detector for optimal coverage and high patient throughput | |
Manual Movement | Vertical travel with electric magnet brake |
Vertical Travel Range (central beam- floor) | Max. : 188 cm (74 in.) Min. : 40 cm (15.7 in.) |
Power Requirement | 24 VDC, 1A |
-Specifications subject to change without notice.