Technical Specifications
Model | CSD-100 |
---|---|
Measure principle |
ISE Method |
Measuring Range |
0~100 to 0~500 ppb |
Display |
0.001 ppb |
Accuracy |
±0.1 ppb or ±5% of reading |
Response Time |
2 min |
Temperature Compensation |
Automatic |
Calibration |
Automatic by standard solution |
Sample Pressure |
1 ~ 5 Bar |
Sample Temperature |
5 ~ 45 ℃ |
Sample Flow |
100~200 ml / min |
Sample Connections |
Inlet 6mm / Outlet 10mm |
Reagent Consumption |
90 days (A/1L, STD/2L*2) |
Data Output |
4 ~ 20 mA 2 ea, Alarm 2 ea |
AUX ports |
DI(2), DO(2), Option (AI, AO) |
Dimensions |
400W X 500H X 200D (Flush Type) |
Degree of Protection |
IP65( NEMA 4x): Electronics, IP55(NEMA 4): Wet part |
Power |
100~240 VAC (50/60 Hz) |