Biometric signal measuring machine manufacturer, Patient monitor, Blood pressure measuring instrument, OEM products, SpO2 sensor and more
KSM-04
SpO2 Module
ㆍ Measuring range : 100% ~ 35%
ㆍ Pulse rate range : 0 ~ 254 bpm
ㆍ Accuracy : SpO2 : 100% ~ 90% : ±1%
89% ~ 70% : ±2%
79% ~ 65% : ±3%
ㆍ below 65%: not specified
ㆍ Pulse : 30 ~ 255bpm ±1bpm
ㆍ SpO2 value & pulse value averaging : 8 pulse beats
KNM-05
NIBP Module
ㆍ Dimension : applied front page
68×128×32mm(Width ×Length ×Height)
ㆍ Weight : approximately 180g
ㆍ Supply Power Voltage : DC 7Volt ±10%, 500mA
ㆍ Consumption Power : 3.2 Watt at Operating Motor
0.35 Watt at Standby
ㆍ Measuring Range : 50 ∼ 280mmHg Systolic, Diastolic and Mean
30 ∼ 300bpm Pulse Rate
ㆍ Measuring Mode : Adult/Neonate
Manual/Automatic
ㆍ Protection : Double Overpressure Protection(Meet IEC Regulation)
Overpressure 300mmHg + 20mmHg at Adult Mode
150mmHg + 15mmHg at Neonate Mode
Hardware Protection - Use of 2 Pressure Sensor
Use of 2 Solenoid Valve
Source·Sink Driver
Software Protection
- Detect Measuring Over Time
- Detect Malfunction of Controller
ㆍ Manometer Mode : 0 ∼ 300mmHg ±3mmHg
ㆍ Operating Temperature : 0 to +50 ℃
ㆍ Operating Humidity : 10% ∼ 95% (noncondensing)
ㆍ Storage Temperature : -20℃ to 60 ℃
ㆍ Measuring Method : Oscilometric
KBC-B04
ECG Module
ECG
ㆍCMRR (Common Mode Rejection Ratio) : ≥60dB
ㆍ Input Impedance ≥10MΩ
ㆍ ECG Channel 1 or 2
ㆍ Electrodes 3 wires : RA, LA, LL
5 wires : RA, LA, LL, RL, V1(Chest)
ㆍ Leads Select 3 Leads : I, II, III
7 Leads : I, II, III, aVR, aVL, aVF, V1
ㆍ Frequency Bandwidth 0.5~40Hz(at Filter )
ㆍ Gain Selection 0.5, 1, 2, 4 mV/cm
ㆍ Heart Rate Pulse Range 30~300 bpm
ㆍ QRS Indication Available
ㆍ Heart Rate Accuracy ±1bpm or 2%
ㆍ Isolation Breakdown Voltage >4000V AC 60 Sec 60Hz
ㆍ Patient Leakage Current ≤100uA
ㆍ Notch Filter 50/60 Hz
ㆍ Defibrillator / ESIS Protected
Respiration
ㆍ Excitation Current < 500uA
ㆍ Excitation Frequency 50kHz
ㆍ Measuring Method Thoracic Impedance Detection between RA and LA of ECG Electrode
ㆍ Measuring Range 2~99 bpm
ㆍ Accuracy ±1bpm or 2%
ㆍ Sensitivity 0.2~4Ω/cm
Temperature
ㆍ Measuring Range 15.0~45.0℃
ㆍ Probe Compatible YSI probe and Exacon
ㆍ Scale Celsius Degrees
ㆍ Accuracy ±0.1℃
Power
ㆍ Direct Current Voltage 6.5~8.0V
ㆍ Consumption Current Maximum 190mA at 7.5V
ㆍ Dimension 120ⅹ68ⅹ16mm(WⅹDⅹH)