Infrared thermometer manufacturer, .Gas measuring device, Gas detector, Smoke detector, Air quality measuring device and more
IR-80 non-contact infrared thermometer measures the infrared wavelength emitted from the target and converts it to standard current signal output (4~20mA) or RS-485 communication signal output. It can measure from 500 ℃ to maximum 1700℃ in the distance of 80:1 D:S (Distance to Spot). Emissivity is 0.10 ~ 0.99 adjustable. Two built-in laser pointers can aim at the target. IR-80 is composed of sensor & controller. The measurement temperature displays on LCD in controller. Various parameters are adjustable in controller.
[ Applications ]
Aluminum, Chrome, Copper, Metal, Magnesium, Oxide-nickel, Platinum, Gold, Silver, Oxided-Titanium, Zinc, Tin, Steel, Oxided- Steel, Oxided-Brass
[ Specification ]
Measurement Range | 500~1700℃ |
Device | InGaAs, Silicon |
Accuracy | ±1% / full scale or 2℃ |
Repeatability | ±1% of reading |
Field of View(D:S) | 80:1 |
Optical spectrum wave | 1~1.6㎛ |
Responsive Time | 100msec or below |
Emissive rate | 0.10~0.99 |
Analog Output | 4~20mA, 1~5V(option) |
Communication output signal | RS-485 communication signal |
Relay Output | 2 step- Relay Contact (AL1/AL2) |
Power | DC 12~24V(Max 100mA) |
Ambient temperature(no water cooling) | 0~70℃ |
Temperature Resolution | 0.1℃ |
Operating Relative Humidity | 5~90% |
Operating Ambient Temperature | -30~85℃ |
Waterproof | IP65,NEMA 4 |
Laser pointer | 630~670nm(red) |
Dimensions | Ø36 ×117.5(L) |
Signal Cable | 4 wire shield type |
Casing material | Aluminum Alloy |
Weight | 850g |
Cable length | 3m(standard), other(option) |
[ Code Number IR-80-□-□-□ ]
Model | Description |
IR-80 | |
Code A | Temperature Range |
1 | 500~1500℃ |
2 | 500~1700℃ |
3 | 800~2200℃(Silicon sensor) |
4 | 150~700℃ (Extend InGaAs 2.6um) |
Z | Other |
Code B | Output |
M | 0~20mA |
N | 4~20mA |
V | Voltage Output(1~5V) |
Code C | Cable Length |
1 | 3m Cable |
Z | Other |
[OPTICAL FIELD OF VIEW (D:S = 80:1)]