Multispectra, Multifiber, Multichannel Imaging spectrometer with 8-16-32 Simultaneous UV-NIR Spectra
HORIBA designs and manufactures a large range of custom OEM fiber-coupled, multi-spectra spectrometers capable of simultaneous measurement of up to 32 fiber channels and sequential measurement up to 96. Utilizing a concentric imaging spectrometer with a broadband spectral range and a customized fiber bundle, the HORIBA OEM multifiber spectrometers provide excellent imaging quality with minimum crosstalk.
The PoliSpectra® MultiTrack systems feature high throughput/ sensitivity and a 2D Scientific back-illuminated CMOS sensor which can be configured with 8,16 or 32 fiber input channels for simultaneous acquisition of UV-NIR spectra.
With an optional nitrogen purge, the system can be configured for low UV coverage. Additionally the high QE sCMOS sensor and an integrated order sorting filter allow wavelength coverage beyond 1 μm.
For OEM Industrial Applications
ㆍ Multitrack Spectroscopy
ㆍ Fluorescence, Emission and Reflectance
ㆍ Some examples from semiconductor applications to life sciences:
- Quality Control
- Light Source Calibration Monitoring
- Plasma Monitoring
- Blood/DNA Multisampling
ㆍ Simultaneous measurement up to 32 fiber channels with minimal crosstalk
ㆍ Extended UV capability below 200 nm
ㆍ High spectral and spatial resolution combined with high throughput
ㆍ High speed and low noise sCMOS 2048 pixels sensor
ㆍ On-board processing capability using powerful dual-core ARM cortex and 2 GB memory
ㆍ USB 3 Interface
ㆍ Stable and robust
Image of 8-channel Fiber Bundle Illuminated with Hg and Ar Lamp
Throughput Differences and Crosstalk between Input Fiber Channels
Scientific CMOS Back-Illuminated Sensor Quantum Efficiency
Thermal stability study cycling from -20ºC to 50ºC
Hyperspectral Imaging Distortions: Keystone and Smile
OEM Spectrometer Selection Guide
8-channel Spectrometer Specifications* | |
---|---|
Spectral Coverage | 185 - 1050 nm (extension to low UV with N2 purge) Tunable at factory between these ranges: 185-920 nm and 280-1050 nm |
Spectrometer Input | 8 channel, φ200 μm core fibers with φ220 μm clad, separated by a φ220 μm dead fiber |
Spectral Resolution | 1.25 nm typical (22 μm slit width) and up, depending on the slit size |
Spectral Dispersion | 34.6 nm/mm; 0.38 nm/pixel |
Crosstalk (Adjacent Channel)* | < 0.1% (measured @ 600 nm) |
Crosstalk (Non-adjacent Channel)* | < 0.1% (measured @ 600 nm) |
Focal Length | 116 mm |
F/# | F/2.3 |
Smile and Keystone | 1.5 pixel (typical) |
Stray Light*** | < 0.1% (typical) |
Wavelength Accuracy | 0.1 nm (using 8 channel wavelength calibration) |
Software | LabVIEW™ acquisition software for initial evaluation (DLLs provided for software integration) |
16-channel and 32-channel Spectrometer Specifications | |
---|---|
Spectral Coverage | 185 - 1050 nm (configurable based on customer’s need) |
Spectrometer Input | 16 channel, φ100 μm core fibers with φ110 μm clad 32 channel, φ70 μm core fibers with φ77 μm clad |
Spectral Resolution | 1 nm and up depending on fiber core size |
Spectral Dispersion | 34.6 nm/mm; 0.38 nm/pixel |
Focal Length | 116 mm |
F/# | F/2.3 |
Smile and Keystone | 1.5 pixel (typical) |
Stray Light | < 0.1% (typical) |
Wavelength Accuracy | 0.1 nm (using 32 channel wavelength calibration) |
Specifications for Scientific Back-Illuminated CMOS Image Sensor with High Speed Electronics | ||
---|---|---|
Detector Model |
sCMOS BI sensor with electronic rolling shutter (all the parameters are specified at normal room temperature of 25º C unless otherwise noted) |
|
Sensor Format**** | 2048 x 512 | |
CMOS Pixel Size | 11 μm x 11 μm | |
CMOS Height | 5.6 mm | |
CMOS QE | The peak QE is 77% at 250 nm and 95% at 560 nm | |
Thermoelectric Cooling | Uncooled | |
Frame Rate | 94 fps (HDR mode), 188 fps (STD mode) on 2D sCMOS sensor | |
Full Well | 80 ke¯ typical (HDR mode) | |
Readout Noise | 1.6 e¯ typical, 2.0 e¯ maximum | |
Digitization | 16-bit (HDR mode), 12-bit (High Sensitivity mode) | |
Dynamic Range | 93.9 dB typical in HDR mode | |
Non-linearity (measured on each system) | 1.8% (maximum) in HDR mode | <0.9% (maximum) in High Sensitivity mode |
Communication | USB 3 | |
Environmental Conditions |
Operating temperature 15°C to 40°C ambient; Relative humidity <70% (non-condensing); Storage temperature -25°C to 45°C |
|
Power requirements AC/DC power supply (provided) |
90-264 VAC, 47–63 Hz |
* Specifications, form factor, and spectrometer cover subject to change without notice. No LabVIEW license is needed to run our acquisition software.
** Crosstalk (to adjacent channels): the % of signal leaking to the adjacent channels while one channel is on and others are off.
*** Stray light using BP Filter: Baseline light level, outside the band, divided by BP peak (unsaturated) value.
Mechanical Drawings